GC APPLICATIONS | ENVIRONMENTAL

Semivolatiles Polycyclic Aromatic Hydrocarbons Rt®-PAH

1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16.

naphthalene acenaphthylene acenaphthene fluorene phenanthrene anthracene fluoranthene pyrene benzo(a)anthracene chrysene benzo(b)fluoranthene benzo(k)fluoranthene benzo(a)pyrene indeno(1,2,3-c,d)pyrene dibenzo(a,h)anthracene benzo(ghi)perylene

Complete separation of PAHs in less than 13 min.

GC_EV00894

Column: Sample: Inj.: Inj. temp.: Carrier gas: Oven temp.: Detector:

Rt®-PAH, 12m, 0.25mm ID, 0.15µm (cat.# 19733) 16 component EPA Method 610 PAH standard (20ng/µl of each component in dichloromethane) 1.0µL split (split ratio 10:1) 225°C helium, 110kPa column head pressure 80°C to 220°C @ 40ºC/min., 220°C to 285°C @ 8°C/min. (hold 5 min.) FID @ 290°C

Chromatogram courtesy of J&K Scientific.