692
Hydrocarbons Hydrocarbons (C30-C110) MXT®-1HT Sim Dist C80
Bleed Profile MXT®-1HT Sim Dist
C70
70
50
C60
60
10
C50
C110
C30
C40
20
C44
30
C100
430°C
40
0 0
10
20
30
40
min.
50
40
30
20
GC_PC00527
min.
GC_PC00530
Column:
MXT®-1HT Sim Dist, 5m, 0.53mm ID, 0.10µm (cat.# 70100) Sample: 0.2µL Polywax® 1000 standard (cat.# 36227) Solvent: carbon disulfide Oven temp.: 40°C to 430°C @ 10°C/min. (hold 30 min.) Injector: on-column (track oven) Carrier gas: helium (constant pressure) Head pressure: 1.0psi Linear velocity: 60cm/sec. Flow rate: 7.8mL/min. Det.: FID @ 430°C Make-up gas flow: 40cc/min.
Column:
MXT®-1HT Sim Dist, 5m, 0.53mm ID, 0.10µm (cat.# 70100) Oven temp.: 40°C to 430°C @ 10°C/min. (hold 30 min.) Injector: on-column (track oven) Carrier gas: helium (constant pressure) Head pressure: 1.0psi Linear velocity: 60cm/sec. Flow rate: 7.8mL/min. Det.: FID @ 430°C Make-up gas flow: 40cc/min.
Hydrocarbons (C44-C100) MXT®-1HT Sim Dist
Hydrocarbons (C10-C44) MXT®-1HT Sim Dist
C10 C9
(fast GC)
GC_PC00531
5
10
15
20
GC_PC00529
Column: Sample: Solvent: Oven temp.: Injector: Carrier gas: Head pressure: Linear velocity: Flow rate: Det.: Make up gas flow:
MXT®-1HT Sim Dist, 5m, 0.53mm ID, 0.10µm (cat.# 70100) 0.2µL injection of Polywax® 1000 carbon disulfide 40°C to 430°C @ 60°C/min. (hold 25 min.) on-column (track oven) helium (constant pressure) 1.0psi 60cm/sec. 7.8mL/min. FID @ 430°C 40cc/min.
Phone: 800-356-1688 or 814-353-1300 www.restek.com
Column: Sample: Solvent: Oven temp.: Injector: Carrier gas: Head pressure: Linear velocity: Flow rate: Det.: Make-up gas flow:
C44
C40
C36
C32
C28
C24
C20
C18
C16
C14
C12
C11
G C A P P L I C AT I O N S | P E T R O L E U M & P E T R O C H E M I C A L S
C90
pA
MXT®-1HT Sim Dist, 5m, 0.53mm ID, 0.10µm (cat.# 70100) 0.2µL hydrocarbon standard (cat.# 31222) carbon disulfide 40°C to 430°C @ 10°C/min. (hold 30 min.) on-column (track oven) helium (constant pressure) 1.0psi 60cm/sec. 7.8mL/min. FID @ 430°C 40cc/min.
25
min.