MEA Layout HighDense IR MEA

HighDenseMEA HD30/10iR-ITO High Density Microelectrode Array with Internal Reference Electrode Technical Specifications...

0 downloads 90 Views 119KB Size
HighDenseMEA HD30/10iR-ITO High Density Microelectrode Array with Internal Reference Electrode

Technical Specifications: HighDenseMEA Temperature compartibility Dimension (W x D x H) Thickness (region of electrodes) Base material Contact pads and track material Electrode diameter Interelectrode distance (centre to centre) Distance between electrode fields Electrode height Electrode type Isolation type Electrode impedance Electrode layout grid Number of electrodes Reference electrodes

0 - 125 °C 49 mm x 49 mm x 1 mm 50 μm Glass Indium tin oxide (ITO) 10 μm 30 μm 500 μm Planar Titanium nitride (TiN) Silicone nitride 500 nm (PEVCD) Approximately 250 - 400 k 2x(5x6) 60 with internal reference electrode

MC_Rack: Source Layout in Data Source Setup MCS Channel map

MEA perfusion chamber

Multi Channel Systems MCS GmbH Aspenhaustrasse 21 72770 Reutlingen Germany

2 dim. (MEA) HighDenseMEA.cmp HighDenseMEA_L.cmp HighDenseMEA_R.cmp (w/o) (gr) (pr) (pr-T)

Without ring Glass ring: ID +/- 20 mm, OD 24 mm, height 6 / 12 mm Plastic ring without thread: ID 26 mm, OD 30 mm, height 6 / 3 mm Plastic ring with thread: ID 25 mm, OD 30 mm, height 6 / 15 mm

Fon +49-7121-9 09 25- 0 Fax +49-7121-9 09 25-11

© 2010 Multi Channel Systems MCS GmbH

[email protected] www.multichannelsystems.com

Product information is subject to change without notice.

HighDenseMEA HD30/10iR-ITO

Electrode Layout

MEA1060 pins 33

21

Electrode #

C3L

A3L

32

31

A4L A5L

44

43

41

42

B4L

B5L

C4L

C5L

52

51

53

54

61

62

71

63

C1R C2R B1R B2R A1R A2R C3R A3R

22 B3L

B3R 72

12 A2L

Left electrode field

A4R 82

Right electrode field

A5R 73

23 A1L A1L

A2L

A3L

A4L

A5L

Electrode #

A1R

A2R

A3R

A4R

A5R

13 B2L

23

12

33

32

31

MEA1060 pins

61

62

63

82

73

B4R 83

34 B1L

B1L

B2L

B3L

B4L

B5L

B1R

B2R

B3R

B4R

B5R

B5R 64

34

13

22

44

43

53

54

72

64

74

C1L

C2L

C3L

C4L

C5L

C1R

C2R

C3R

C4R

C5R

14

24

21

41

42

52

51

71

83

84

D1R

D2R

D3R

D4R

D5R

57

58

78

75

85

24 C2L 14 C1L

25 D2L

L R

D2L

D3L

D4L

D5L

15

25

28

48

47

E1L

E2L

E3L

E4L

E5L

E1R

E2R

E3R

E4R

E5R

35

16

27

45

46

56

55

77

86

65

F1L

F2L

F3L

F4L

F5L

F1R

F2R

F3R

F4R

F5R

26

17

36

37

38

68

67

66

87

76

REF

15 D1L

35 E1L 16 E2L

C4R 74 C5R 84 D5R 85 D4R 75 E5R 65

26 F1L

500 μm

F5R 76

30 μm 10 μm

17 F2L

E4R 86

F4R 87

27 E3L

E3R 77

F3L

D3L

F4L

F5L

E4L

E5L

36

28

37

38

45

46

D4L D5L D1R D2R E1R 48

47

57

58

56

E2R

F1R

F2R D3R F3R

55

68

67

78

66

The first letter of the electrode number code refers to the row number, the digit is the column number, and the second letter refers to the electrode field (left or right) of the HighDenseMEA. The specified MEA1060 pin numbers are the channel numbers that are used in the MC_Rack program. The electrode D1 of the left electrode field, connected to channel 15 in MC_Rack is missing. It is replaced by a big internal reference electrode.

Multi Channel Systems MCS GmbH Aspenhaustrasse 21 72770 Reutlingen Germany

Fon +49-7121-9 09 25- 0 Fax +49-7121-9 09 25-11

© 2010 Multi Channel Systems MCS GmbH

[email protected] www.multichannelsystems.com

Product information is subject to change without notice.